Safety SubCulture Unit Quick Guide
Note: This document is a product use reference guide. Please consult the package insert for the complete Instructions for Use.
Insert A) Remove SCU from B) Hold bottle with one C) Hold SCU in place
package and position hand. With other hand, with one hand and
SCU tip at center of press down on SCU using other hand, pull
disinfected bottle flanges to pierce bottle up on white filter cap
stopper. top. to remove.
Note: The SCU will not
sit flush with the
stopper.
Drop A) Position SCU B) Tilt bottle to C) Replace white filter
sampling channel dispense drop. cap when finished
Method close to media plate dispensing drops.
or slide.
Syringe A) Aseptically remove B) Insert male end of C) Upright bottle and
the sampling channel. syringe into female luer syringe. Pull up on
Method For example, place an of SCU. Invert bottle syringe plunger to
alcohol wipe over and syringe and clear any remaining
insert and pull up with withdraw required culture in the SCU.
finger tips. volume into syringe.
D) Replace white filter Remove A) Hold bottle with one
cap as described hand and with other
Σελίδα 1
D) Replace white filter Remove A) Hold bottle with one
cap as described hand and with other
above in the Drop and hand twist and pull up
Method. Dispense Discard on SCU to remove.
sample from syringe.
SCU Discard into appropriate
bio-hazard container.
PTM-ITLAP-004 REV01
ITL Corporation
t: +(1) 888 411 2851
f: +(1) 703 435 6717 Sharps Safety & Best Practice
sales@itlus.com
www.itlcorporation.com Copyright 2009 ITL Corporation. All rights reserved.
Σελίδα 2
Safety SubCulture Unit Tips & Tricks
General Guidelines
When dispensing to a
Tilt bottle to smaller angle Prepare to upright
series of slides or other
(approx. 30-45 degrees) bottle quickly to cut off
media, dispense drops
above horizontal for better the drop size and rate.
sequentially without
control of drop size and
uprighting the bottle
speed.
between drops.
Note: Gaseous samples
tend to flow more quickly.
Resin Media
To reduce potential Increase the angle of If resin obstructs the SCU
clogging, turn the bottle the bottle to dispense tip inside the bottle, tilt the
horizontal allowing the drops. wide end of the bottle
resin to settle along the downward to clear the resin
side of the bottle prior to from the tip. Allow the resin
dispensing drops. to settle along the side of
the bottle prior to
dispensing additional drops.
Charcoal Media
To reduce potential Clogging may be Clogging may also be
clogging, tilt the bottle cleared by gently cleared by replacing the
approx. 30-45 degrees pressing the base of SCU filter cap and gently
above horizontal. Do the SCU against the tapping the base of the
not invert the bottle, as septum while the bottle bottle on a counter.
this may increase is at a slight angle.
Σελίδα 2
the SCU against the tapping the base of the
not invert the bottle, as septum while the bottle bottle on a counter.
this may increase is at a slight angle.
clogging. An alternative is to use
the syringe method.