Performance of three ready-to-use bottles for grinding orthopaedic samples Olivier Dauwalder1, Camille Kolenda1, Cécile Eymard1, Marine Jawhar1, Cécile Batailler2, Christophe Gaillard2, Frédéric Laurent1, Céline Dupieux1 1 Bactériologie, Hospices Civils de Lyon, France 2 Chirurgie orthopédique, Hospices Civils de Lyon, France Céline DUPIEUX Bacteriology Department - Institute of Infectious Agents - Croix Rousse Hospital Rhône-Alpes Reference Center for Complex Osteoarticular Infections celine.dupieux@chu-lyon.fr

Page 2

INTRODUCTION ▪ ▪ Diagnosis of essential IOA by culture (antibiogram) ▪ Grinding of solid samples required before seeding ▪ Objective: optimise sensitivity / limit contamination ▪ Medium: sampling media allowing direct grinding without handling the sample ▪ OBJECTIVE = to compare the analytical performance of two new ready-to-use vials • ProbeAx Evolution (PAx ; Axonlab A.G.) • TISSUtainer (TIt ; Diagante) au flacon Ultra-Turrax® UTX-001 (UTx ; Labelians)

Page 3

SUPPORTS EVALUES Ultra-Turrax® UTX-001 TISSUtainer ProbeAx Evolution (UTx ; Labelians) (TIt ; Diagante) (PAx ; Axonlab) Grinding medium Sterile water quality Quality preservation Quality saline solution BM 2 mL environment BM, 10 mL BM 5 mL Grinding method Ceramic block + stainless Steel beeds x10 Steel beeds x4 steel metal grater Grinder Ultra-Turrax® (IKA) MM400 (Retsch®) SpinAX (AxonLab A.G.) Grinding time 1 min / 1 sample 2,5 min / 2-8 samples 1 min / 1 sample Marking CE (old regulations 98/79/ DM-DIV sterile class A, DM-DIV sterile class A, CE) CE IVDR CE IVDR

Page 4

EQUIPMENT AND METHODS ▪ Prospective study ▪ 2 orthopaedic surgery units at the Hospices Civils de Lyon (HCL) ▪ 15 patients with strong suspicion of IOA • Diagnostic samples taken for routine care (RC) in Ultra-Turrax tubes • Among these samples, 2 taken in triplicate and introduced into the operating room in the 3 tubes • In the laboratory: - Ultra-Turrax tubes processed according to the RC flow - ProbeAx and TISSUtainer bottles processed for the study according to the availability of the technicians in charge (pre-analytical storage at +4°C if waiting; 1-3 days)

Page 5

INTERPRETATION OF RESULTS ▪ Culture reading: quantification/identification of colonies obtained ▪ Definitions: • Species responsible for AIO: if found ≥ 2 times on SC or study swabs for potentially contaminating bacteria and ≥ 1 time for non-classically contaminating bacteria • Contamination if only 1 positive sample with a species that is not very virulent in flora or the environment ▪ Comparaison : • Nombre d’espèces bactériennes retrouvées avec chaque poudrier • Référence : résultats globaux des pvts en Ultra-Turrax du SC • Pour chaque espèce retrouvée par ≥ 2 types de poudriers, comparaison semi-quantitative du nombre de colonies obtenues (catégories > 50 UFC, 26-50, 11-25, 6-10, 1-5, uniquement bouillon)

Page 6

RESULTS ▪ The 3 to 7 UTx vials produced for the SC of the 15 patients were recovered. • 7 infections monomicrobial • 5 infections polymicrobial • no microorganisms for 3 patients

Page 7

RESULTATS ▪ Pour les 3 patients avec prélèvements stériles Patient Ultra-Turrax SC TISSUTainer ProbeAx Evo 2 2/2 stériles 2/2 stériles 2/2 stériles 5 2/2 stériles 2/2 stériles 2/2 stériles 1/2 stérile 14 2/2 stériles 2/2 stériles et 1/2 contaminé * * 1 colonie Cutibacterium acnes sur SCS14

Page 8

RESULTS ▪ For the 7 monomicrobial infections: detected in 6 out of 7 cases (86%) in at least 1 of the 2 samples in the study, across all vials compared (UTx, PAx, TIt). Pvts SC Ultra-Turrax Additional species in Patient Microorganisme TISSUTainer ProbeAx globaux SC the study 3 Finegoldia magna 4/5 1/2 1/2 2/2 4 Staphylococcus aureus 5/5 2/2 1/2 1/2 Corynebacterium striatum (TIt et 7 Pseudomonas aeruginosa 5/5 2/2 2/2 2/2 PAx) ; Proteus mirabilis et Streptococcus constellatus (PAx) 9 Cutibacterium acnes 5/5 2/2 2/2 2/2 10 Staphylococcus aureus 4/4 2/2 2/2 2/2 13 Cutibacterium acnes 2/5 0/2 0/2 0/2 1 contaminant en TIt et 1 en PAx 15 Cutibacterium acnes 5/5 2/2 2/2 2/2

Page 9

RESULTS ▪ For the 5 polymicrobial infections (16 bacterial species in SC): 14 (87.5%) detected in UTx bottles, 11 (69%) in PAx and 12 (75%) in TIt. Pvts SC UltraTurrax Patient Microorganisme TISSUTainer ProbeAx Additional species in the study globaux SC 1 Escherichia coli 5/7 2/2 2/2 2/2 1 Staphylococcus aureus 1/7 0/2 0/2 0/2 6 Enterobacter cloacae 5/5 2/2 2/2 2/2 6 Enterococcus faecium 5/5 2/2 2/2 1/2 8 Pasteurella canis 2/3 1/2 2/2 2/2 Neisseria spp en plus 8 Fusobacterium spp 2/3 2/2 2/2 2/2 avec PAx et TIt 8 Parvimonas micra 1/3 0/2 0/2 0/2 11 Proteus mirabilis 5/5 2/2 2/2 2/2 1 contaminant en PAx 11 Staphylococcus aureus 2/6 2/2 0/2 0/2 12 Escherichia coli 5/5 2/2 2/2 2/2 12 Staphylococcus aureus 5/5 2/2 2/2 2/2 Staphylococcus epidermidis 12 Enterococcus casseliflavus 4/5 2/2 2/2 2/2 en plus avec PAx et TIt 12 Acinetobacter nosocomialis 5/5 2/2 1/2 2/2 1 contaminant en PAx 12 Lysinibacillus fusiformis 2/5 1/2 0/2 0/2 12 Bacillus thuringiensis 5/5 2/2 2/2 2/2 12 Streptococcus mitis-oralis 4/5 1/2 2/2 0/2

Page 10

RESULTS - SUMMARY Overall SC results (3 Results for 2 UTx Results for 2 TIt Results for 2 PAx to 7 UTx) 3 patients with stériles stériles stériles sterile samples - - 6 detected by ≥ 1 of the - - 6 detected by ≥ 1 of - - 6 detected by ≥ 1 of 7 IOA monomicrobial samples included the samples included the samples included - - 1 not detected - - 1 not detected - - 1 not detected 5 IOA polymicrobial (16 5/5 detected 5/5 detected 5/5 detected species) (14 species) (12 species) (11 species) Additional species: Additional species : - contaminations: 3 - contaminations: 2 (C. (non-aureus staphylococci) acnes, Kocuria) - pathogens: 5 - pathogens: 3 (C. striatum, Neisseria spp, (C. striatum, Neisseria spp, P. mirabilis, S. epidermidis, S. epidermidis) S. constellatus) → Ultimately, the results from the new bottles were strictly consistent or detected more species than the SC UTx in 76.7% of cases for both PAx and TIt.

Page 11

RESULTS ▪ Pre-analytical storage at +4°C for up to 3 days of PAx and TIt bottles did not prevent the detection of fastidious species (Pasteurella, Neisseria, anaerobes). ▪ The number of colonies obtained on the culture media was not significantly different despite different grinding volumes depending on the flasks (UTx: 2 mL; PAx: 5 mL; TIt: 10 mL). Semi-quantitative categorization: > 50 CFU, 26-50, 11-25, 6-10, 1-5, broth only

Page 12

DISCUSSION-CONCLUSION ▪ Prospective study, samples introduced into the new powder containers as soon as they leave the operating room ▪ small number of patients -> larger prospective studies needed ▪ Demonstrates at least equivalent analytical performance between the new TISSUtainer and ProbeAx Evo vials vs. Ultra-Turrax. ▪ Other practical criteria should be taken into account. : • bag and bottle holder, • impact resistance (pneumatic), • noise (UTx > TIt > PAx), • space requirement (TIt > UTx et PAx), • number of samples processed at the same time (TIt > Utx et PAx) • or volume of available mulch (TIt > PAx > UTx)

Page 13

ACKNOWLEDGEMENTS ▪ To the technicians and biologists of the bacteriology department at the Institute of Infectious Agents, Hospices Civils de Lyon Cécile EYMARD Olivier DAUWALDER Marine JAWHAR Camille KOLENDA ▪ To orthopedic surgeons who included patients Pr Cécile BATAILLER Dr Christophe GAILLARD